MAGNETIC THICKNESS GAUGE OF DUAL-LAYER COATINGS MTGD-1


      The device measures two wide ranges of thicknesses of ingredients of dual-layer chrome-nickel coatings (chrome on nickel, and nickel under chrome) on non-ferromagnetic and weak-magnetic basis.
      The device is completed with two magneto-dynamic transducers. Magnet parameters of one of them are chosen for providing of high resolution for chrome thickness; the resolution is constant in the given range of variation of nickel thickness, and in this range one can test chrome thickness independently of nickel thickness. Magnet parameters of other transducer are chosen for providing of high resolution for nickel thickness, minimal resolution for chrome thickness and approaching of nickel magnetization in the volume of informative area to saturation magnetization, therefore one can test thickness of nickel under chrome with elimination or minimization of error that is conditioned by nickel structure.

Specifications

Measurement of thickness of non-ferromagnetic (chrome, etc.) coating on nickel coating:
     - measurement range of thickness of non-ferromagnetic coating, micrometers

0-150

     - admissible variation range of thickness of nickel coating, micrometers

200-700

Measurement of thickness of nickel coating under non-ferromagnetic (chrome, etc.) coating:
     - measurement range of thickness of nickel coating, micrometers

0-700

     - admissible variation range of thickness of non-ferromagnetic coating, micrometers

0-150

Limits of basic error, micrometers:
     - when thickness of non-ferromagnetic coating on nickel coating with thickness 200-700 micrometers is been measuring

(5 + 5% of measurable thickness)

     - when thickness of nickel coating under non-ferromagnetic coating with thickness up to 150 micrometers is been measuring

(5 + 10% of measurable thickness)

Maximum number of memorized zero adjustments

891

Duration of one measurement, seconds, no more than

1

Power supply

4 alkaline batteries or accumulators (AAA type)

Interval of automatic shutdown after measurement stoppage, minutes, no more than

6

Overall dimensions, millimeters, no more than:
    - the electronic module

150x80x30

    - the transducer

Ø19x55

Weight of the device, grams, no more than

300

State calibration, warranty and after-sales service

Magnetic thickness gauge of dual-layer coatings MTGD-1 is entered in the State registry of measuring gauges of the Russian Federation
(registration number 50930-12, certificate 47796)

Phone/fax (017) 284-23-07; E-mail: Lukhvich@iaph.bas-net.by